User's Guide _______________________________________________________________________
112 __________________________________________________________________ M210762EN-A
>fcrh
RH : 11.25 Ref1 ?
RH : 11.25 Ref1 ?
RH : 11.25 Ref1 ?
RH : 11.24 Ref1 ?
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
9. Now the device is waiting for the wet end reference. Insert the
probe in the wet end reference condition (30 ... 75 %RH). Press
any key when ready.
10. Let the probe stabilize for about 30 minutes. You can follow the
stabilization by pressing Enter.
11. When stabilized, type the high end reference value after the
question mark and press Enter.
>fcrh
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
RH : 75.45 Ref2 ? c
RH : 75.57 Ref2 ? c
RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>
12. The OK indicates that the adjustment has succeeded and the new
calibration coefficients are calculated and stored. Enter the
adjustment information (date and text) to the memory of the
transmitter, see section
Adjustment Information on page 109.
13. Press the ADJ button on the motherboard to close the
adjustment mode.
14. Take the probe out of the reference conditions and replace the
filter.