Chapter 6 __________________________________________________________________Settings
VAISALA _______________________________________________________________________ 27
Device Information
The basic information about the indicator and the probe is found as
follows:
1. Open the MENU: press OPEN
2. Select Settings, press .
3. Select Device information, press SHOW.
4. The first display gives the information on the MI70 indicator.
Press MORE to get the information on the probe. Press OK
and EXIT to return to the basic display.
Reverting to Factory Settings
You can revert to factory settings to clear all changed settings and
recorded data on the indicator. Reverting to factory settings does not
affect probe calibration.
1. Open the MENU: press OPEN.
2. Select Settings, press .
3. Select Factory settings, press REVERT. Press YES to confirm the
reverting.
4. The power turns off automatically. When switching on again,
the factory settings are reverted. After reverting to the factory
settings, you must set the language, date and time again. See
Chapter 3.
Probe type information
MI70 indicator
information