1. Take out the screw covering the calibration button (in the probe
handle).
2. Press the calibration button. For pressing, use tool with thin and
sharp point, such as small screwdriver. When pressing the
button, indicator turns to adjusting mode.
3. Select T, press SELECT.
4. Now the adjustment mode is on, press ADJUST to select the
adjustment method: 1-point adjustment or 2-point adjustment.
1-Point Adjustment
Follow the steps from 1 to 4 and continue as follows:
5. Select 1-point adjustment, press SELECT.
6. Set the probe to a reference temperature. You can follow the
stabilization from the GRAPH display. Press READY when the
reading is stabilized in the reference.
7. Give the reference temperature value by using the arrow
buttons. Press OK.
8. Confirm the adjustment, press YES. (By pressing NO you
return to adjustment mode display and no changes are made).
9. Calibration is carried out. Press BACK to exit the adjustment
mode and EXIT to return to the basic display.