User's Guide ______________________________________________________________________
30 __________________________________________________________________ M210297EN-E
5. Press SELECT.
6. To set the recording duration, select Duration, press SET.
7. Select the recording duration (1 min, 5 min, 15 min, 30 min, 1 h, 3
h, 12 h, 24 h, 7 d, 30 d, memory full) with arrow buttons and
press SELECT.
8. Start recording: select Start/Stop recording, press START. If you
chose 'Memory full', you see the maximum recording time on
the display. You can also delete files in order to empty the
memory. Press START again to accept maximum recording
time.
9. You can switch the HM70 off during recording to save battery.
Display message tells you that recording continues undisturbed
even when power is off. If the indicator is switched off during
recording, the progress bar is shown on the display every
10 seconds (all the time, if charger is connected). This bar shows
the amount of recorded data.
CAUTION
Do not disconnect the probe when the data recording is on,
even if the indicator is off. This may cause loss of recorded
data.
Stopping Recording
1. To stop recording press Record, select Record data and press ,
select Start/Stop recording and press STOP.
2. Now you can go and see the recorded file by selecting
SHOW.
You can save individual measurement data points with Hold/Save
function described in page 23.
Viewing Recorded Data
1. Open the MENU: press OPEN .
2. Select Recording/Viewing, press .
3. Select View recorded data, press .
Comentarios a estos manuales