Chapter 11 __________________________________________________ Analog Output Connection
VAISALA _______________________________________________________________________ 59
CHAPTER 11
ANALOG OUTPUT CONNECTION
Selecting and Scaling the Analog
Output
To get analog measurement data, you need the analog output signal
cable, see list of accessories on page 77. One voltage signal channel
0...1.0V can be scaled for the selected quantity.
1. Connect the analog output signal cable connector to the indicator
base connector. Connect the screw terminal block as follows:
Brown wire: common wire (-)
Yellow-green wire: signal (+)
2. Open the MENU: press OPEN.
3. Select Functions, press .
4. Select Analog output, press .