HUMIDITY
556
HUMIDITY
9
WHEN YOU NEED IT RIGHT, RIGHT NOW, CALL KELE.
kele.com888-397-5353 USA
March 2014
INFRARED DEW POINT & GRAINS TRANSMITTER
9002
5.63
(14.30)
3.25
(8.26)
1.37
(3.48)
in
(cm)
MOUNTING
DIMENSIONS
ORDERING INFORMATION
MODEL DESCRIPTION
9002 Dew point and grains 0-10 VDC/4-20 mA transmitter and SPDT contact
Tamper Set Screw
+
L1N
NONC
Input
AC
COM
Mounting Bracket
24 VAC
Use either or both
outputs below
0-10V Out
Signal Common
4-20 mA Out
1238765 412
WIRING
clear
set
enter
Mode
Button
Clear
Button
Arrow
Button
Enter
Button
CONTROLS
SETUP
FACTORY OUTPUT SETTINGS
Dew point, 0° to 80°F (-22° to 25°C) @ sea level
Moisture = 0-140 grains per lb of dry air @ sea level
Relay on @ 50°F (10°C), off @ 59°F (15°C)
CHANGING SETTINGS
1. Press mode and clear at the same time for five seconds
to enter the adjust mode.
2. Press mode key to display item to change:
elevation, dew point, grains, or relay.
3. Use arrow keys to select ranges.
4. Press enter to confirm selection.
5. Press clear to finish and return to normal display.
RELATED PRODUCTS PAGE
T1508 Duct-mount accessory enclosure
T1551 OSA-mount accessory enclosure
691-K0A Control transformer, 120:24 VAC, 40 VA, Class 2 979