Vaisala HUMICAP HMW90 SERIES Especificaciones Pagina 7

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 74
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 6
501
HUMIDITY
HUMIDITY
9
WE MAKE IT EASY.
kele.com 888-397-5353 USA
March 2014
in
(cm)
2.6
(6.7)
5.3 (13.5)
3.5
(9.0)
6.4 (16.3)
2.6
(6.7)
5.3 (13.5)
3.5
(9.0)
Plastic enclosure
Metal enclosure
1
Serial data output
2
3
4
5
6
7
8
Analog output
Supply
RxD/B-
TxD/A+
GNDSer
Out2
Out1
GND
GND
V+
Power Supply
8-35 VDC
12-30VAC
4-20 mA +
4-20 mA +
-
MULTIFUNCTION HUMIDITY/TEMPERATURE TRANSMITTER
E31 SERIES
Accuracy
Humidity ±1.3% RH, <90% RH, 5° to 104°F
(-15° to 40°C)
Temperature ±0.27°F (±0.14°C) ±0.06
Supply Voltage
0-5/10 VDC:
8-35 VDC, 40 mA; 12-30 VAC,
1.0 VA
4- 20 mA: 8-35 VDC, 80 mA; 12-30 VAC,
1.9 VA
Signal Output Field selectable: 4-20 mA, 0-20 mA,
(500 max load), 0-5 V, or 0-10 V,
Communication Serial interface RS232C
Measurement Range
Humidity 0 to 100% RH
Temperature -40° to 120°F (-40° to 48.9°C)
Sensing Technology
Humidity Capacitive
Temperature Pt1000 RTD, 385 curve
Sensor Protection Stainless steel sintered filter
Display Type Optional, two line, alphanumeric
LCD, English units
Display
Dewpoint -40° to 120°F (-40° to 48.9°C)
Enthalpy 0-64 Btu/lb (0 to 400 kJ/kg)
Wetbulb 0° to 100°F (-17.8° to 37.8°C)
Operating Temperature
Display: -40° to 140°F
(-40° to 60°C)
No Display -4° to 122°F (-20° to 50°C)
Enclosure NEMA 4, IP65
Wiring Terminations Screw terminals
Warranty 1 year
DESCRIPTION
The E+E E31 Series Multifunction Humidity/Temperature
Transmitters offer precise measurement of humidity and
temperature with field selectable analog outputs for both
parameters. The unit can also be ordered with enthalpy, dew
point, and wet bulb. Two-point calibration can be performed
using push-buttons on the circuit board-no need for
dismounting which makes it ideal for regulatory environments.
Built-in LEDs indicate transmitter status and errors. An
optional display is available for local indication of measured
values and minimum and maximum values.
FEATURES
Accuracy of ±1.3% RH
Derived values for dew point, enthalpy, and wet bulb
Simple calibration
Stainless steel sintered filter
Optional display available
Dual output signals
RS232 output
SPECIFICATIONS
WIRING
DIMENSIONS
Vista de pagina 6
1 2 3 4 5 6 7 8 9 10 11 12 ... 73 74

Comentarios a estos manuales

Sin comentarios