User's Guide _______________________________________________________________________
10 ____________________________________________________________________M210498en-A
CHAPTER 4
INSTALLING THE PROBE
Selecting a place for the probe
Select a place which gives a true picture of the process. Oil should
circulate freely around the sensor; a rapid oil flow is recommended. Install
the probe directly into the circulation system and not into the oil reservoir
because of deposition.
It is recommended that the sensor head is installed directly in the process
through the ball valve assembly. When the ball valve assembly is used, the
pipe does not have to be emptied or shut down for installation or removal
of the probe. Install the sensor head transversely against the direction of
the process flow. Avoid installing into a bend in the pipe.
NOTE
Take care not to damage the pipe of the probe. If the
pipe is damaged, the probe head is less tight and will
not go through the clasp nut. Make sure that the
filter is tightly fastened to protect the sensors.
Mounting the probe for pressurized pipelines/oils
Due to the sliding fit, the MMP78 is easy to install into and remove from
the pressurized process. The probe is especially suitable for the
measurements in pipelines. The maximum installation pressure is 10 bar
and the maximum operation pressure is 20 bar.