User's Guide _______________________________________________________________________
20 ____________________________________________________________________M210498en-A
Feeding the coefficients to the MI70
memory
You can save in the memory specific coefficients for 4 different oils. As
shipped from factory the average coefficients are set for the number 1 type
(H
2
O
oil
: 1). Add your own coefficients as follows:
1. Open the MENU; press OPEN
2. Select SETTINGS and press .
3. Select MEASUREMENT SETTINGS and press .
4. If you want to keep the average coefficients under identification
number 1 (set in the factory) press
SET and give number 2 for the
second oil type.
5. Select the A value (A:0.0000) and press SET.
6. Set the value A by using using buttons, press OK.
7. Select the B value (B:0.0000) and press SET.
8. Set the value B by using using buttons, press OK.
9. You may give coefficients for 4 different oils. Indentify the oil by
numbering them (1...4). Select the first row (H
2
O
oil
) and set the
identification number. Then feed the coefficients as shown above.
10. Press EXIT to return to the basic display.
11. Remember always select the correct oil type before measurements.
Oil
identification
number 1-4