Vaisala HMP228 Guía de usuario Pagina 38

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 57
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 37
User's Guide _______________________________________________________________________
36 ____________________________________________________________________M210498en-A
Relative humidity adjustment by
using a calibrated reference probe
Follow the steps from 1 to 12 and continue as follows:
13. Ensure that the probes are located in the same conditions.
14. Wait until the readings have stabilized. If the difference between the
humidity readings is less than 2 % RH, there is no immediate need
for adjustment.
15. Adjust the transmitter reading to correspond a reading of the
reference probe by pressing ADJUST.
16. Select To same as RH
I/II
. Press SELECT.
17. Confirm by pressing YES
18. Press BACK, EXIT, EXIT to return to the basic display.
1-point relative humidity adjustment
by using a calibrator
You can use Vaisala humidity calibrator HMK15 to achieve the reference
humidities. MI70 indicator is working as a communicator in the calibration
procedure.
Follow the steps from 1 to 12 and continue as follows:
13. Insert the transmitter probe to the reference humidity
14. Press ADJUST
15. Select 1-point adjustment. press SELECT
Vista de pagina 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 56 57

Comentarios a estos manuales

Sin comentarios