User's Guide ______________________________________________________________________
98 __________________________________________________________________ M210566EN-J
Formatting Serial Line Message
NOTE
Instead of using the FTIME, FDATE and FST commands described in
this section, you can use the FORM command with modifiers TIME,
DATE, and STAT. See section FORM on page 101.
FTIME and FDATE
FTIME and FDATE commands will enable/disable output of time and
date to the serial line. To add time to R and SEND outputs enter:
FTIME [x]<cr>
To add date to R and SEND outputs enter:
FDATE [x]<cr>
where
x = ON or OFF
Examples:
>send
RH= 98.4 %RH T= 31.0 'C
>ftime on
Form. time : ON
>send
03:47:59 RH= 98.4 %RH T= 31.0 'C
>fdate on
Form. date : ON
>send
2004-07-05 03:48:03 RH= 98.4 %RH T= 31.0 'C
>