User's Guide ______________________________________________________________________
118 _________________________________________________________________ M210566EN-J
Example:
>play 3 2007-05-05 00:00:00 2007-05-06 00:00:00
RH (12 min intervals) 2007-05-05 00:00:00 121
Date Time trend min max
yyyy-mm-dd hh:mm:ss %RH %RH %RH
2007-05-05 00:00:00 19.16 18.99 19.33
2007-05-05 00:12:00 19.30 19.09 19.55
2007-05-05 00:24:00 20.01 19.28 21.17
2007-05-05 00:36:00 21.21 20.98 21.44
2007-05-05 00:48:00 19.57 17.72 21.11
2007-05-05 01:00:00 19.09 18.62 19.84
...
You can use the <ESC> key to interrupt the output listing.
NOTE
Output of large amounts of recorded data can result in huge data files
and take a long time, up to several days for the entire memory of the
data logger at 10 second resolution. To make it easier to process the data
it is recommended to select the largest suitable data interval, and to
specify the start and end times carefully.
Deleting the Recorded Files
You can delete the recorded data files using the keypad/display, or the
DELETE command on the serial line. The deletion is always done for all
data; you cannot delete individual files.
Note that the transmitter automatically overwrites the old data when the
memory is full, so manual deletion of the recorded files is not necessary
in normal use.
To delete the data files using the keypad/display:
1. Press any of the arrow buttons to open the Main Menu.
2. Select System by pressing the ► arrow button.
3. Select Clear graph memories by pressing the CLEAR button.
Press the YES button to confirm the selection.
CAUTION
This function clears the entire data history of the transmitter, including
all graphs and the content of the optional data logger module.