Use the HELP command to list the currently available commands. The
available commands are determined by the device configuration and
installed options.
Example:
>help
? ACAL ADDR AERR ALSEL
ASEL CDATE CLOSE CON CRH
CT CTEXT DATE DELETE DIR
DSEL DSEND ECHO ERRS FCRH
FILT FORM HELP INTV ITEST
LIGHT LOCK MODBUS MODS PLAY
PRES R RESET SCOM SDELAY
SEND SERI SMODE SYSTEM TEST
TIME UNDELETE UNIT VERS XPRES
>
ERRS
Use the ERRS command to display transmitter error messages, see Table
32 on page 152.
Example (no active errors):
>errs
No errors
>
Example (showing active error):
>errs
Error: E2 Humidity sensor open circuit.
>
VERS
Use the VERS command to display software version information.
Example:
>vers
HMT330 / 5.10
>
Resetting Transmitter Using Serial Line
RESET
This command resets the device. The user port switches to start-up output
mode selected with command SMODE.