DMT345 and DMT346 Dewpoint Transmitters
for High Temperature Applications
The Vaisala DRYCAP
®
Dewpoint
Transmitters DMT345 and
DMT346 are designed for humidity
measurement in industrial drying
applications with particularly high
temperatures.
Both transmitters incorporate the
Vaisala DRYCAP
®
sensor, which is
accurate, reliable, and stable. The
sensor withstands condensation
and is immune to particulate
contamination, oil vapor and most
chemicals. The DRYCAP
®
sensor
stands out for its swift response time
and rapid recovery after getting wet.
Measure humidity directly
in hot processes
The DMT345 and DMT346 are
constructed for direct measurement
in hot processes. Therefore, there
is no need for sampling systems
and trace heating. As a result,
high accuracy and constancy are
maintained.
The Vaisala DRYCAP
®
Dewpoint Transmitters DMT345 and DMT346 are designed
to measure and control humidity especially in dry environments with high
temperatures.
The accuracy and stability of the
DMT345 and the DMT346 are due to
the unique auto-calibration function,
patented by Vaisala. This auto-
calibration makes the transmitter
perform a calibration and adjustment
by itself while the measured process
is running. If the measurement
accuracy is not conrmed, corrections
are made automatically. The
procedure is so quick and corrections
are so minor that it will go unnoticed.
This ensures low maintenance
and high performance. In normal
conditions, it is recommended to have
a traceable calibration performed
once a year.
DMT345, Accurate in hot
and dry environments
The DMT345 is designed for accurate
humidity measurement in hot and
dry conditions. This model provides
unmatched dry end measurement
accuracy in temperatures up to 140 °C,
however the DMT345 can operate
safely in temperatures up to 180 °C.
The stainless steel probe is especially
designed for high temperatures and
has an optional installation ange that
allows an adjustable installation depth
and therefore a precise positioning.
Features/Benefits
The DMT345 measures
humidity in temperatures up to
180 °C (356 °F)
The DMT346 measures
humidity in temperatures up to
350 °C (+662 °F)
Dewpoint accuracy ±2 °C
(±3.6 °F)
Vaisala DRYCAP® Sensor
provides accurate, reliable
measurement with excellent
long-term stability and fast
response
Withstands condensation
Unique auto-calibration feature
Optional local display with
keypad, mains power supply
module and alarm relays
NIST traceable calibration
(certificate included)
Analog outputs, WLAN/LAN
The large and clear display allows the
user to check data at a glance.
Technical data
Measured variables DMT345
DEWPOINT DMT345
Sensor Vaisala DRYCAP
®
180S
Measurement range -40 ... +100 °C (-40 ... +212 °F) Td
Accuracy ±2°C (±3.6 °F) Td
See the accuracy graph below
Response time 63 % [90 %] flow rate 1 l/min and 1 bar pressure
from dry to wet 5s [10 s]
from wet to dry including auto-calibration 45s [5 min]
Dewpoint accuracy vs. measurement conditions
DMT346, Reliable in very hot processes
When process temperatures range between 140 °C to
350 °C, the DMT346 provides the best measurement
performance.
The DMT346 comes with a cooling set as a standard
feature. The cooling effect may be regulated by adding the
cooling ns, or removing them from the set for the best
measurement performance.
The cooling system operates without moving parts,
additional power or cooling utilities, thereby eliminating
the risk of sensor damage due to a mechanical cooling
failure.
Additionally, sensor warming minimizes the risk of
condensing on the sensor. In low humidity the combination
of auto-calibration and DRYCAP
®
ensures accurate
measurement.
Graphical display
The DMT345 and DMT346 transmitters can be ordered
with a large numerical and graphical display, which allows
the user to clearly monitor operational data, measurement
trends and up to one-year measurement history.
The optional data logger with real-time clock makes it
possible to generate more than four years of the measured
history and zoom in on any desired time or time frame.
Versatile outputs and (wireless) data
collection
The transmitter can be connected to a network with an
optional (W)LAN interface, which enables a (wireless)
Ethernet connection.
For serial interface also the USB connection, RS232 and
RS485 can be used. Additionally an alarm relay option is
available.
The transmitter can have up to three analog outputs.
Galvanic isolation of supply power and analog outputs
are also offered. The recorded measurement data can be
viewed on the display or transferred to a PC with Microsoft
Windows
®
software.
Units are delivered installation-ready and meet ROHS
requirements.
TEMPERATURE DMT345
Measurement range 0 ... +180 °C (+32 ... +356°F)
with sensor warming upper range limited by humidity
(at 80% RH warming is switched on and T
reading not actual process Temperature)
Accuracy ±0.4 °C at 100 °C
Temperature sensor Pt 100 IEC 751 1/3 class B
RELATIVE HUMIDITY DMT345
Measurement range 0 ... 100 % RH
with sensor warming 0 ... 80 % RH
Accuracy
below 10 % RH ±10 % of reading
above 10 % RH ±1,5% RH + 1,5 % of reading
MIXING RATIO DMT345
Measurement range (typical) 0 ... 1000 g/kg (0 ... 7000 gr/lbs)
Accuracy (typical) ±12% of reading