HMP260 SERIES
HMP260-U017en-1.5 Appendix 5: Error messages
The LEDs indicate these errors as follows:
¡¡¤¤ no frequency at all
¡¡¤¡ a frequency missing from the RH channel
¡¡¡¤ a frequency missing from the T channel
The second error type indicates erroneous y-values (used in internal calculations):
Display: Serial bus:
E51 T y-value out of range
E53 U1 y-value out of range
E54 U2 y-value out of range
The LEDs indicate these errors as follows:
¤¡¡¡ RH channel y-value out of range
¡¤¡¡ T channel y-value out of range