HMP260 SERIES
Operating Manual HMP260-U017en-1.5
36 1997-11-25
5.4.2.6 Selecting the output units
UNIT x <cr>
x = m(etric units)
n(on-metric units)
metric non-metric
RH %RH %RH
C °F
Td °C °F
ag/m
3
gr/ft
3
x g/kg gr/lb
Tw °C °F
5.4.2.7 Setting the transmitter address
ADDR aa <cr>
aa = address (0...99)
Example:
transmitter is given address 99
ADDR <cr>
Address : 2 ? 99 <cr>
5.4.2.8 Resetting the transmitter
RESET <cr>
5.4.3 Operating the transmitter via the serial bus
5.4.3.1 Setting the serial interface
SMODE xxxx<cr>
xxxx = STOP, RUN or POLL
In STOP mode: measurements output only by command, all commands can be
used
In RUN mode: outputting automatically, only command S can be used
Vista de pagina 41