HMP260 SERIES
Operating Manual HMP260-U017en-1.5
18 1997-11-25
Fig. 4.5 Connection of cables
When the serial bus has been connected between the PC and the transmitter,
the PC is switched on. When using a PC, a terminal emulation programme
(e.g. Procomm Plus, Datastorm or Windows terminal) is started.
The factory settings for data transfer are:
4800 baud
even parity
7 data bits
1 stop bit
full duplex
NOTE
When the serial bus settings are changed, the transmit-
ter has to be reset before the new settings become ef-
fective.
The processor does not allow the following combinations:
no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP260 programme will change the number of stop bits to 2
even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the programme changes the number of stop bits to 1
Refer to the manuals of the PC and the terminal emulation programme when
giving serial settings.
2
5
3
3
7
2
3
7
2
TX
GND
RX
TX
GND
RX
TX
GND
RX
HMP 2
3
D9S
D25S
D25P
PC
TXD
RXD
TXD
TXD
RXD
RXD
4
6
7
8
5
6
8
20
TERMINAL