HMP260 SERIES
HMP260-U017en-1.5 Operating Manual
1997-11-25 33
5.4.1.2 Selecting and scaling the analogue output quantities
Disconnect the security lock jumper!
ASEL xxx yyy <cr>
xxx = channel 1's quantity
yyy = channel 2's quantity (RH, T, Td, Abs, Mix or Tw)
Example:
relative humidity selected on channel 1 and temperature on channel 2
ASEL RH T <cr>
Ch1 (RH) lo 0.000 %RH ? <cr>
Ch1 (RH) hi 100.000 %RH ? <cr>
Ch2 (T ) lo -40.000 'C ? <cr>
Ch2 (T ) hi +160.000 'C ? <cr>
5.4.1.3 Scaling the analogue outputs
Disconnect the security lock jumper!
ASCL <cr>
Example:
relative humidity is scaled on the range of 0...100 %RH and temperature -
40...+160 °C
ASCL <cr>
Ch1 (RH) lo 0.000 %RH ? <cr>
Ch1 (RH) hi 100.000 %RH ? <cr>
Ch2 (T ) lo 0.000 'C ? -40 <cr>
Ch2 (T ) hi 100.000 'C ? 160 <cr>