HMP260 SERIES
Operating Manual HMP260-U017en-1.5
12 1997-11-25
push back the braid and press it between the two brass disks to
achieve a full 360° grounding; the fold between the disks should have
the same diameter as the brass disks
secure the braid with a shielding tube
insert the wires into the transmitter housing through the gland.
tighten the nut
connect the wires into the screw terminals and fasten a cable tie
around the wires
(If the cable diameter is less
than 7mm, use a shrinking
tube or an adhesive tape)
cable
braid
flexible wires 0.5 mm
(AWG 20), stranded wires
recommended
2
D = Ø 7...10
3
brass
disks
25
rubber
ring
nut
brass disks
140
165
shielding tube
braid
transmitter housing
cable tie
gland
brass disks
rubber ring
nut