APPENDIX 7 ________________________________________________________________________
VAISALA ________________________________________________________________________ 137
When the alarm outputs have been set, you can activate the outputs
with commands ALARM 1 ON (channel 1) or ALARM 2 ON
(channel 2). Note that you can also activate the outputs when giving
other parameters with command ALARM.
Examples of activating the outputs:
with command ALARM:
>alarm 1 Td HI -20 5
Ch1 Td HI ON -20.00 5.00 ’C
Ch2 RH HI ON 100.00 0.00 %RH
>
with commands ALARM 1 ON and ALARM 2 ON:
>alarm 1 ON
Ch1 Td HI ON -20.00 5.00 ’C
Ch2 T HI OFF 30.00 5.00 ’C
>alarm 2 ON
Ch1 Td HI ON -20.00 5.00 ’C
Ch2 T HI ON 30.00 5.00 ’C
>