CHAPTER 6_______________________________________________________________ CALIBRATION
VAISALA _________________________________________________________________________ 57
Press any key and enter the second point value and press <cr>.
RH : xx.x Ref2 ? yy.y <cr>
The stabilization of the sensor can be monitored here as well by
entering c <cr> instead of the reference value.
Using display/keypad commands
Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize.
Remove the filter cap on the transmitter.
Disconnect the security lock jumper
before turning the transmitter on!
Place the sensor head in the calibration hole of the LiCl bottle (dry
end reference) in the humidity calibrator.
Select &DOL in the main menu and then 5+; select 1RW FKDQJHG
and then two-point calibration 5+  SRLQW FDO. Change the first
point reading with the arrow keys to correspond the reference
humidity and press ENT; pressing an arrow once changes the
reading by 0.05 %RH.
Place the sensor head in the calibration hole of the NaCl bottle
(wet end reference) in the humidity calibrator.
Wait for 10 minutes.
If necessary, change the second point reading with the arrow keys
and press ENT.
Using LED commands
Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize.
Remove the filter cap on the transmitter.
Disconnect the security lock jumper
before turning the transmitter on!