OPERATING MANUAL_______________________________________________________________
54 _____________________________________________________________________ U263EN-1.4
Example: relative humidity calibration is performed at transmitter
2 which is in POLL state
>OPEN 2 <cr>
opens the line to transmitter 2
>CRH <cr>
calibration started
...
>CLOSE <cr> line closed