APPENDIX 1 SERIAL COMMANDS__________________________________________________________
VAISALA _________________________________________________________________________ 93
Any text can be written in the command and it appears in the output. For ex-
ample:
RH: \UUU.U\ T: \+TT.TT\\r RH: 54.0 T: +25 <cr>
The format can be deleted by giving \ as a parameter:
>FORM \<cr> Note. only one space before \ and none after
or
>FORM <cr>
"xxx...xxx"
?\<cr>
An example of a format suitable for use in Microsoft Excel spreadsheets:
>FORM <cr>
"xxx...xxx"
?\UUU.U\\t\TTT.T\\t\DDD.D\\t\\r\n <cr>
The output is then:
47.4<tab> 22.4 <tab> 10.6 <tab> <cr><lf>
FTIME Adding time to output
FTIME xxx <cr>
xxx = ON or OFF
When FTIME is activated, the current time is output at the beginning of the
output line. The time is set with command TIME. After RESET or power on
the current time is 00:00:00.
Activating the time output
>ftime on
Form. time : ON
>intv 5 s
setting the output interval
Output intrv. : 5 s
>r
09:31:13 RH= 19.4 %RH T= 26.0 ’C
09:31:18 RH= 19.4 %RH T= 26.0 ’C
09:31:23 RH= 19.8 %RH T= 26.0 ’C
09:31:28 RH= 19.6 %RH T= 26.0 ’C
09:31:33 RH= 19.5 %RH T= 26.0 ’C
09:31:38 RH= 19.5 %RH T= 26.0 ’C
...