Chapter 4 _________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 103
Example:
>fst on
Form. status : ON
>send
w 0 Tdf= 3.9 'C Td= 3.9 'C Tdfa= 3.9 'C Tda= 3.9 'C
H2O= 8045 ppmV x= 5.0 g/kg RH= 26.8 %RH a= 5.9 g/m3
aNTP= 6.4 g/m3 T= 24.1 'C dT= 20.3 'C
>purge
Purge started, press any key to abort...
>send
S 120.3 Tdf= 3.8 'C Td= 3.8 'C Tdfa= 3.8 'C Tda= 3.8 'C
H2O= 7999 ppmV x= 5.0 g/kg RH= 12.5 %RH a= 5.6 g/m3
aNTP= 6.4 g/m3 T= 37.4 'C dT= 33.6 'C
>
where the state of the probe is indicated by the following letters
and
values:
w...xxx
=
Sensor warming
where xxx
=
Sensor heat power
(0 when not warming)
H...xxx
=
Purge
where xxx
=
Sensor temperature (ºC)
S...xxx
=
Sensor cooling
after purge
where xxx
=
Sensor temperature (ºC)
A...xxx
=
AutoCal
where xxx
=
Sensor temperature (ºC)
General Settings
Changing the Quantities and Units
To change quantities and units use serial commands or the optional
display/keypad. See Table 2 on page 18 for available quantities and
Table 3 on page 18 for optional quantities.
NOTE
Only the quantities
selected when ordering the device can be selected as
an output quantity.
Using Display/Keypad
Use display/keypad to select the display output quantities.
1. Press any of the arrow buttons to open the Main Menu.
2. Press the arrow button to select Display.
3. Press arrow button to select Quantities.
4. Select the quantity by using the ▲▼ arrow buttons. Confirm the
selection by pressing SELECT. You can select 1 ... 4 display
quantities at a time.
5. Press EXIT to return to the basic display.