Chapter 3 ________________________________________________________________ Installation
VAISALA ________________________________________________________________________ 31
Grounding the Cables
Ground the screen of the electrical cable properly to achieve the best
possible EMC performance.
0605-027
Figure 16 Grounding the Screen of Electrical Cable
Refer to Figure 16 above when performing the procedure below.
1. Cut back outer sheath to desired length.
2. Cut back screen braiding or screen foil to dimension X.
3. Push the domed cap nut (item 1) and the seal insert with contact
socket of the gland (item 2+3) onto the cable as shown in the
diagram
4. Bend over the screen braiding or screen foil by about 90º
(item 4).
5. Push the seal insert with the contact socket of the gland
(item 2+3) up to the screen braiding or screen foil.
6. Mount the lower part (item 5) on the housing
7. Push the seal with the contact socket of the gland (item 2+3)
into the lower part (item 5).
8. Fasten the domed cap nut (item 1) onto the lower part (item 5).