Vaisala DRYCAP DMT340 SERIES Guía de usuario Pagina 112

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 191
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 111
User's Guide _______________________________________________________________________
110 __________________________________________________________________ M210704EN-G
FILT
Use the serial line command FILT to set the filtering level.
FILT [level] [extfactor]<cr>
where
level
=
Filtering level. The options are:
OFF
(no filtering)
ON (standard filtering
, enabled by default.
A
pproximately 13 s moving average)
EXT (extended filtering, approximately 1 min average
by
default but configurable by changing the extfactor)
extfactor
=
Weighted average factor for extended filtering level.
Range 0 ... 1, default 0.030.
In extended filtering mode,
the output is calculated
based
on the following formula:
[(
new result * extfactor) + (old result * (1.0 - extfactor))]
This
means that with extfactor value 1 the transmitter
will take only the latest measure
ment into account, but
with
extfactor value 0.1 a new output is a combination of
the previous output (90%) and the latest measurement
(10%).
Example (enable extended filtering):
>filt ext
Filter : EXT
>
Vista de pagina 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 190 191

Comentarios a estos manuales

Sin comentarios