Analog Output Settings
The analog outputs are set in the factory according to the order form. In
case you want to change the settings, follow these instructions. See
section Third Analog Output on page 56.
Changing the Output Mode and Range
Both output channels have their own dip switch module with 8 switches,
see the position in Figure 2 on page 21 (dip switches for analog output
settings).
1. Select the current/voltage output, switch ON either of the switches,
1 or 2.
2. Select the range,switch ON one of the switches from 3 to 7.
0503-045
Figure 68 Current/Voltage Switches of Output Modules
The numbers refer to Figure 68 above:
1
=
Current/voltage selection output switches (from 1 to 2)
2
=
Current/voltage range selection switches (from 3 to 7) in analog
output 1 and 2
3
=
Switches for service use only. Keep in OFF position always.
NOTE
Only set one mode and one range switch to ON for one output
channel.
The rest of the switches must remain OFF.