User's Guide _______________________________________________________________________
166 __________________________________________________________________ M210704EN-G
ppmv (dry)
Measurement range (typical) 10 ... 2500 ppm
Accuracy (at + 20 °C, 1 bar) 1 ppm + 20% of reading
Other measurement parameters available (depends on model)
mixing ratio, absolute humidity, pressure dewpoint calculated to 1 bar,
temperature difference (T-Td), water vapor pressure
Operating Environment
Mechanical durability up to +180 °C (+356 °F)
for transmitter body -40 ... +60 °C (-40 ... +140 °F)
with display 0 ... +60 °C (32 ... +140 °F)
Storage temperature range
without display -55+80 °C (-67+176 °F)
with display -40+80 °C (-40+176 °F)
Pressure and temperature range for probes see probe specifications
Measured gases non corrosive
Electromagnetic compatibility EN61326-1: Electrical
equipment for measurement,
control, and laboratory use
EMC requirements for use in
industrial locations
Inputs and Outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC 50/60 Hz
Default start-up time after sensor Purge and Autocal About 6 min
Start-up time after power-up 3 s
Power consumption at 20 °C (Uin 24VDC)
RS-232 max 25 mA
Uout 2 x 0 ... 1V / 0 ... 5V / 0 ... 10V max 25 mA
Iout 2 x 0 ... 20 mA max 60 mA
display and backlight + 20 mA
during sensor purge + 110 mA max
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V
Accuracy of analog outputs at 20 °C ± 0.05 % full scale
Temperature dependence of the analog outputs ± 0.005 %/°C full scale
External loads
current ouputs RL < 500 ohm
0 ... 1V output RL > 2 kohm
0 ... 5V and 0 ... 10V outputs RL > 10 kohm
Wire size 0.5 ... 2.5 mm
2
(AWG 20 ... 14)
stranded wires recommended
Digital outputs RS-232
RS-422/485 (optional)
LAN (optional)
WLAN (optional)
Protocols ASCII commands
Modbus RTU
Modbus TCP