User's Guide _______________________________________________________________________
38 ___________________________________________________________________ M210704EN-G
NOTE
-tightening the nut after detachment the nut must be tightened
without increased effort.
6. Clean and grease the tightening cone of the fitting screw after every
tenth detachment. Change the sealing washer every time the fitting
screw is detached. Use high-vacuum grease (for example Dow
Corning) or a similar grease.
0503-033
Figure 23 Cleaning of the Tightening Cone
The numbers refer to Figure 23 above:
CAUTION
In pressurized processes it is essential to tighten the supporting nuts and
screws very carefully to prevent
loosening of the probe by the action of
NOTE
When the DMT340 is installed in a process with a pressure differing
from the selected operating pressure at the time of ordering, please enter
the pressure value of the process into the transmitter m
emory. You can
in the following ways:
using the display/keypad option
using the buttons on the motherboard
using the serial line
using the Modbus protocol
Pressure Compensation Setting on page 106. If you are
Modbus, see section Configuration Registers on page 185.