User's Guide _______________________________________________________________________
164 __________________________________________________________________ M210704EN-G
Adjusting Analog Outputs using Serial
Line
ACAL
Use the serial line to perform the analog output adjustment. Use the
ACAL command and enter the multimeter reading.
ACAL<cr>
Example (current outputs):
>ACAL
Ch1 I1 (mA) ? 2.046
Ch1 I2 (mA) ? 18.087
Ch2 I1 (mA) ? 2.036
Ch2 I2 (mA) ? 18.071
>
Press the adjustment button on the motherboard inside the transmitter
again to disable the adjustment function.