User's Guide _______________________________________________________________________
30 ___________________________________________________________________ M210934EN-D
Format Measurement Output
Use the serial line command FORM to change the format or select
certain parameters for the output commands SEND and R.
FORM [x]<cr>
where
x
=
Formatter string
Formatter string consists of parameters and modifiers. The MMT162
measures the following parameters:
- water activity (aw)
- temperature (T) (metric unit: ºC, non metric unit:ºF)
- ppm for transformer oil only (H
2
O)
When selecting the parameter, use its abbreviation (for example, aw).
The modifiers are presented in Table 6 on page 31.
Example (configure transmitter to output water activity and temperature):
form "aw=" 6.4 aw #t "t=" 6.2 t #r#n
OK
send
aw= 0.2644 t= 25.50
Example (configure transmitter to output ppm only):
form "Oil ppm= " h2o " " u3 #r#n
OK
send
Oil ppm= 16.6 ppm
Example (configure transmitter to default output format):
form /
OK
send
aw= 0.087 T= 24.0 'C