Chapter 6 ___________________________________________________ Calibration and Adjustment
VAISALA ________________________________________________________________________ 57
4. Wait for 30 minutes for temperature and humidity to stabilize.
5. Turn on the MI70.
6. Select Display Quantities and units from the menu of the
MI70, and verify that measurements from both ports (I and II)
are available. If the MMT162 is not detected, it may be because it
is already powered up and using non-default serial line settings.
Try the following steps:
a. Turn off the MI70.
b. Disconnect the cable from port I of the MMT162.
c. Turn on the MI70.
7. Select Functions Adjustments (II) from the menu of the MI70
and press Start. Make sure the symbol for port II is shown after the
word "Adjustments", as your reference probe may also be
adjustable using the MI70.
8. The adjustment mode starts. Automatic power off is disabled while
in adjustmend mode; press OK to acknowledge the message.
9. Select the parameter to calibrate and adjust (RH or T) from the list
and press Select. This procedure is applicable for both parameters,
but the following steps will assume you selected RH.
10. Check and adjust the environment settings of the reference probe if
prompted by the MI70.
11. The adjustment mode now shows the value of the selected
parameter from both instruments, and a delta parameter (for
example, ΔRH) that shows the difference between the two. The
difference should be no greater than the combined measurement
uncertainty of the MMT162 transmitter and the reference probe.
Depending on the result, proceed as follows:
- If the MMT162 transmitter is within its accuracy
specification, there is no need to proceed with the adjustment.
Select Back and Exit to leave the adjustment mode.
- If adjustment is required, continue with the next step.
12. Select Adjust, and then select To same as RH (I).
13. The MI70 will ask Do you really want to adjust? Select Yes.
MI70 shows the text Adjustment done and returns to the
adjustment mode after a few seconds. Check the value of the delta
parameter to verify that the adjustment has taken effect.
14. Select Back to return to the parameter selection screen.
15. Select a new parameter for adjustment and proceed as instructed
above, or select Exit to leave the adjustment mode.