Chapter 6 ___________________________________________________ Calibration and Adjustment
VAISALA ________________________________________________________________________ 63
Temperature Adjustment on Serial Line
Required equipment:
- Computer with
- Windows operating system
- Terminal application
- Free USB port
- Driver for Vaisala USB cable installed
- USB serial interface cable (219690)
- One or two temperature references that fulfill the requirements listed
in section Adjustment Points on page 55.
1. If any cable is connected to port I of the MMT162, disconnect it.
2. Remove the filter from the MMT162 and insert the sensors into the
temperature reference. If you are starting a two-point calibration,
start with the lower temperature reference.
3. Wait at least 30 minutes for the temperature to stabilize.
4. Connect the MMT162 to a PC, see section Serial Communication
on page 22. Then open a terminal program.
5. Enter the l command to view the currently active calibration
coefficients. For more information, see section Show User
Calibration Coefficients on page 42.
6. Enter the ct command and press Enter:
ct
7. Type c and press Enter to update the reading and verify that the
reading has stabilized:
T : 22.3106 1. ref ? c
T : 22.3251 1. ref ?
8. When the reading is stable, enter the reference temperature after the
question mark and press Enter:
T : 22.3261 1. ref ? 22.3
Press any key when ready ...