Chapter 5 _________________________________________________________________ Operation
Example 2
In this example the analog output 1 is set to send measurement
data as a current signal, the gain is set to 1 mA/m/s and the offset
to 4 mA.
S aout1_o,0.004
S aout1_g,0.001
S aout1mode,0
Example 3
In this example both analog outputs are disabled to reduce power
consumption.
S aout1mode,3
S aout2mode,7
G Get Parameter
The G command shows the values of the configuration
parameters. You can either display all parameter values or only
certain values.
For a list of available parameters, see Appendix D Configuration
Parameters on page 205.
Get All Parameters
You can use the G command to view all parameter values if you
do not specify any parameters in the command.
G
Get Specified Parameters
You can use the G command to view specified parameter values.
G x
where
x
=
Any parameter in Appendix D Configuration Parameters on
page 205
VAISALA _______________________________________________________________________ 113