User's Guide ______________________________________________________________________
Example 2:
In this example the data message 2 is set to include the following
items:
- SOH
- Checksum calculation start point
- Wind speed, average
- Wind direction, average
- Wind gust speed
- Wind lull speed
- Wind direction minimum
- Wind direction maximum
- Wind direction during the peak speed (wp) occurred
- Checksum calculation end point
- EOT
- Print checksum
- CR
- LF
S
msg2,\01\ss$\ws,\wd,\gu,\lu,\dm,\dx,\w1\se\04\sp\cr\l
f
WMT700 sends the following message that starts with the SOH
character which is excluded from the checksum. The checksum
ends before the EOT character. The checksum (in this case, D8) is
printed after the EOT character.
┌_$02.66,98.21,02.66,02.60,95.68,99.53,99.34_D8<CR><L
F>
The non-printable characters are shown above as ┌ ┘and
<CR><LF>.
122 __________________________________________________________________ M211095EN-E