User's Guide ______________________________________________________________________
The response is the following:
atttn<CR><LF>
where
a
=
Single-digit WMT700 address that corresponds to the
first character of the address value
ttt
=
Period of time, in seconds, after which WMT700 has
the measurement ready
n
=
Number of measurement values WMT700 calculates
and returns as a response to one or more subsequent
aD0! commands
<CR><LF>
=
Terminates the response
Example of the command and response:
1M!
10025<CR><LF>
Interpretation of the example message:
- Time after which the measurement is ready: 2 second
- Number of returned measurement values: 5
To fetch the measurement results, use the aD0! command.