Vaisala HUMICAP HMT310 Guía de usuario Pagina 31

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 105
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 30
Chapter 3 ________________________________________________________________ Installation
VAISALA ________________________________________________________________________ 31
The following two fitting body options are available:
- Fitting Body ISO1/2 solid structure
- Fitting Body NPT1/2 solid structure
Table 4 HMT318 Probe Dimensions
Probe Type
Probe Dimension
Adjustment Range
Standard
178 mm
120 mm
Optional
400 mm
340 mm
0507-025
Figure 12 Sealing of Fitting Body into Process
Tightening the Clasp Nut
1. Adjust the probe to a suitable depth according to the type of
installation.
2. Tighten the clasp nut first manually.
3. Mark the fitting screw and the clasp nut.
4. Tighten the nut a further 50 - 60º (ca. 1/6 turn) with a wrench. If
you have suitable torque wrench, tighten the nut to max 45 ± 5 Nm
(33 ± 4 ft-lbs).
Vista de pagina 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 104 105

Comentarios a estos manuales

Sin comentarios