Vaisala HUMICAP HMT310 Guía de usuario Pagina 75

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 105
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 74
Chapter 6 ___________________________________________________ Calibration and Adjustment
VAISALA ________________________________________________________________________ 75
High End Adjustment
7. After the low end adjustment, insert the probe into the
measurement hole of the high end reference chamber (for example,
NaCl: 75 % RH chamber in the humidity calibrator HMK15, use
the adapter fitting 13.5 mm for HMT314, HMT315, HMT317, and
HMT318). Note that the difference between the two humidity
references must be at least 50% RH.
8. Let the probe stabilize for at least 30 minutes. You can follow the
stabilization by pressing any key, typing C and pressing Enter.
9. When stabilized, type the high-end reference value after the
question mark and press Enter.
>crh
RH : 11.25 Ref1 ? c
RH : 11.24 Ref1 ? c
RH : 11.24 Ref1 ? 11.3
Press any key when ready ...
RH : 75.45 Ref2 ? c
RH : 75.57 Ref2 ? c
RH : 75.55 Ref2 ? c
RH : 75.59 Ref2 ? 75.5
OK
>
OK indicates that the calibration has succeeded.
10. Take the probe out of the reference conditions and replace the
filter. When using the stainless steel filter, take care to tighten the
filter properly, recommended force is 130 Ncm.
11. If needed, record the calibration information (text and date) to the
memory of the transmitter: see Set Calibration Information Text on
page 71 and Set Calibration Date on page 72.
12. Reset the transmitter by issuing the RESET command. The
transmitter returns to normal mode.
Vista de pagina 74
1 2 ... 70 71 72 73 74 75 76 77 78 79 80 ... 104 105

Comentarios a estos manuales

Sin comentarios