Vaisala HUMICAP HMT310 Guía de usuario Pagina 50

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 105
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 49
User's Guide _______________________________________________________________________
50 ___________________________________________________________________ M210619EN-E
Other Commands
Check Transmitter Settings
Syntax: ?<cr>
Syntax: ??<cr>
Use the command ? to check the current transmitter configuration. The
command ?? is similar but can also be used if the transmitter is in POLL
state.
Example (factory default settings):
>?
HMT310 / 1.07
PRB serial nr : A0000000
Calibration : 2004-05-07
Cal. info : Vaisala/HEL
Output units : metric
Pressure : 1013.25 hPa
RS232 settings
Address : 0
Output interval: 0 S
Baud P D S : 4800 E 7 1
Serial mode : STOP
Analog outputs
Ch1 output mode: 0 ...20mA
Ch2 output mode: 0 ...20mA
Ch1 error out : 0.000mA
Ch2 error out : 0.000mA
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
>
Serial Bus Echo
Syntax: ECHO x<cr>
Where
X
=
ON/OFF (default = ON)
The command enables or disables the echo of characters received over
the serial line.
Vista de pagina 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 104 105

Comentarios a estos manuales

Sin comentarios