115
Technical Data
Moisture in Oil
Measured Values
WATER ACTIVITY
Measurement range a
w
0 ... 1
Accuracy (including non-linearity, hysteresis and repeatability)
0 ... 0.9 ±0.02
0.9 ... 1.0 ±0.03
Response time (90%) at +20 °C in still oil
(with stainless steel filter) 10 min.
Sensor HUMICAP
®
180
L
2
Performance
TEMPERATURE
Measurement range
MMT332 -40 ... +180 °C (-40 ... +356 °F)
MMT337 -40 ... +180 °C (-40 ... +356 °F)
MMT338 -40 ... +180 °C (-40 ... +356 °F)
Accuracy at +20 °C (+68 °F) ± 0.2 °C (0.36 °F)
Operating Environment
Operating temperature
for probes same as measurement ranges
for transmitter body -40 ... +60 °C (-40 ... +140 °F)
with display 0 ... +60 °C (+32 ... +140 °F)
Pressure range for probes see probe specifications
Electromagnetic compatibility Complies with EMC standard
EN61326-1, Industrial environment
Note: Transmitter with display test impedance of
40 ohm is used in IEC61000-4-5 (Surge immunity)
Inputs and Outputs
Operating voltage 10 ... 35 VDC, 24 VAC ± 20%
with optional power supply module 100 ... 240 VAC 50/60 Hz
Power consumption @ 20 °C (U
in
24VDC)
RS232 max. 25 mA
U
out
2 x 0...1V / 0...5V / 0...10V max. 25 mA
I
out
2 x 0...20 mA max. 60 mA
display and backlight + 20 mA
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V
Accuracy of analog outputs at 20 °C ± 0.05% full scale
Temperature dependence of the
analog outputs ± 0.005%/°C full scale
External loads
current outputs R
L
< 500 ohm
0 ... 1V output R
L
> 2 kohm
0 ... 5V and 0 ... 10V outputs R
L
> 10 kohm
Max. wire size 0.5 mm
2
(AWG 20) stranded wires recommended
Digital outputs RS232, RS485 (optional)
Protocols ASCII commands, MODBUS RTU
Service connection RS232, USB
Relay outputs 0.5 A, 250 VAC, SPDT, potential-free (optional)
Ethernet interface (optional)
Supported standards 10BASE-T, 100BASE-TX
Connector 8P8C (RJ45)
IPv4 address assignment DHCP (automatic), static
Protocols Telnet, MODBUS TCP/IP
WLAN interface (optional)
Supported standards 802.11b
Antenna connector type RP-SMA
IPv4 address assignment DHCP (automatic), static
Protocols Telnet, MODBUS TCP/IP
Security WEP 64/128, WPA
Authentication / Encryption
Open / no encryption
Open / WEP
WPA Pre-shared key / TKIP
WPA Pre-shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock
Logged parameters max. four with trend/min/max values
Logging interval 10 sec. (fixed)
Max. logging period 4 years, 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years
Display LCD with backlight, graphical
trend display of any parameter
Menu languages English, Chinese, Finnish, French, German,
Japanese, Russian, Spanish, Swedish
Mechanics
Cable bushing M20x1.5 for cable diameter 8 ... 11mm/0.31 ... 0.43"
Conduit fitting 1/2" NPT
Interface cable connector (optional) M12 series 8-pin (male)
option 1 female plug with 5 m (16.4 ft.) black cable
option 2 female plug with screw terminals
USB-RJ45 Serial Connection Cable
(incl. Mi70 Link software)
219685
Probe cable diameter 5.5 mm
Standard probe cable lengths 2 m, 5 m or 10 m
(Additional cable lengths available,
please see order forms for details)
Housing material G-AlSi 10 Mg (DIN 1725)
Housing classification IP 66
IP65 (NEMA4X) with local display
Weight
depending on selected probe, cable and modules 1.0 - 3.0 kgs
Sensor protection Stainless steel grid standard filter/
Stainless steel grid filter for high flow rates (>1 m/s)
Ref. B210953EN-C ©Vaisala 2013
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Please contact us at
www.vaisala.com/requestinfo
www.vaisala.com
Scan the code for
more information
Comentarios a estos manuales