GMP343 Carbon Dioxide Probe for Demanding
Measurements
The GMP343 is available as an open-path diffusion-aspirated model (left) and as a
flow-through model (right).
Features/Benefits
▪
Excellent accuracy and stability
▪
Vaisala CARBOCAP® Sensor,
a silicon-based non-dispersive
infrared (NDIR) sensor
▪
A single-beam, dual-
wavelength CO
2
measurement
with no moving parts
▪
Compensation options
for temperature, pressure,
humidity and oxygen
▪
Low power consumption and
heat emission
▪
Designed for outdoor use
▪
Compact and lightweight
Technical Data
Performance
Measurement range options 0 ... 1000 ppm, 0 ... 2000 ppm,
0 ... 3000 ppm, 0 ... 4000 ppm,
0 ... 5000 ppm, 0 ... 2 %
Accuracy (excluding noise) at 25 °C (77 °F) and 1013 hPa after factory
calibration with 0.5 % accurate gases with different range options
0 ... 1000 ppm ±(3 ppm + 1 % of reading)
0 ... 2000 ppm - 0 ... 2 %* ±(5 ppm + 2 % of reading)
*Accuracy below 200 ppm CO
2
not specified for 2 % range option
Noise (repeatability) at 370 ppm CO
2
with no output averaging ±3 ppm CO
2
with 30 s output averaging ±1 ppm CO
2
TEMPERATURE
Effect on accuracy with temperature compensation:
CO
2
range options 0 ... 1000 ppm 0 ... 2 000 - 5000 ppm
0 ... 2 %
Temperature °C (°F) Accuracy (% of reading)*
+10 ... +40 (+50 ... +104) ±1 ±1 ±2
+40 ... +60 (+104 ... +140) ±2 ±3 ±4
-40 ... +10 (-40 ... +50) ±3 ±3 ±5
* Always at least ±10 ppm CO
2
.
Temperature compensation is performed by an integrated Pt1000 element
The Vaisala CARBOCAP
®
Carbon Dioxide Probe
GMP343 is an accurate and rugged probe-type
instrument for ecological measurements. Typical
applications include CO
2
soil respiration, ambient
CO
2
monitoring, plant growth chambers, and
OEM applications.
The GMP343 can output both numerically filtered
and raw measurement data and it can also
compensate the measurement with an internal
temperature measurement and user-set relative
humidity, pressure and oxygen values.
In combination with an MI70 indicator, the GMP343
provides a tool for accurate in-situ measurement.
The MI70 can be used as a display, communication
and data logging device.
Each GMP343 is calibrated using ±0.5 % accurate
gases at 0 ppm, 200 ppm, 370 ppm, 600 ppm,
1000 ppm, 4000 ppm and 2 %. Calibration is also
done at temperature points of -30 °C, 0 °C, 25 °C
and 50 °C. If needed, the customer can recalibrate
the instrument using the multipoint calibration
(MPC) feature allowing up to 8 user-defined
calibration points.