Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Accuracy (including non-linearity, hysteresis,
and repeatability):
at 0 ... +40 °C (+32 ... +104 °F) ±1.5 %RH (0 ... 90 %RH)
±2.5 %RH (90 ... 100 %RH)
at -10 ... 0 °C and +40 ... +60 °C
(+14 ... +32 °F and +104 ... +140 °F) ±3.0 %RH (0 ... 90 %RH)
±4.0 %RH (90 ... 100 %RH)
Factory calibration uncertainty
at +20 °C (+68 °F) ±1.5 %RH
Humidity measurement response time
(90%) with plastic grid filter 17 s
Stability ±2 %RH over 2 years
Humidity sensor HUMICAP
®
180R
TEMPERATURE
Measurement range -10 ... +60 °C (14 ... 140 °F)
Accuracy over temperature range:
at 0 ... +40 °C (+32 ... +104°F) ±0.2 °C (0.36 °F)
at -10 ... 0 °C and +40 ... +60 °C
(+14 ... 32 °F and +104 ... +140 °F) ±0.4 °C (0.72 °F)
Temperature sensor Pt1000 RTD Class F0.1 IEC 60751
General
Power-up time < 3 s
Batteries 2 × AA sized, 1.5 V
Calculated parameters Td, Tw, a, x, h
Menu languages English, German, French, Finnish, Spanish,
Swedish, Chinese (simplified), Russian, Japanese, Portuguese
Display LCD (140 x 160 pixels)
Operation time typical 100 hours (without backlight)
Operation temperature range -10 ... +60 °C (+14 ... +140 °F)
Storage temperature range -30 ... +70 °C (-22 ... +158 °F)
Electromagnetic compatibility European Union directive
(EMC) EN61326-1 for portable equipment
Accessories and Spare Parts
HMP113 probe for HM40*
HM40 handle and cable HM40HANDLE
Plastic nut for HMP113 attachment (3 pcs) DRW238590SP
NiMH rechargeable batteries (4 pcs) 229247SP
External battery recharger with USB
connection and 2 batteries 229249SP
Case for HM40 (indicator and probe) 235849SP
Plastic grid filter for HMP113 probe DRW236214SP
Plastic grid with membrane filter for HMP113 probe 230727SP
Belt clip (3 pcs) 227710SP
Battery cover (3 pcs) 225688SP
* See the HM40 order form when ordering a spare probe for HM40.
Mechanics
Weight (with alkaline batteries)
With standard probe 230 g
With handle 330 g
Materials
Meter body PC/ABS blend, acrylic display lens
Probe holder PC/ABS blend (grey)
Probe PC/ABS blend (white)
Filter PC (glass reinforced)
Housing classification IP54
Mechanical drop endurance 0.4 m with probe attached
1.0 m without the probe
72
232.5
44
12
155
77.5
24.5
207.5 13.5
36.5
4.4
8.4
68
224
140
12
1200
24.5
70
32
Dimensions
Dimensions in mm
Ref. B211064EN-C ©Vaisala 2013
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Humidity
Vista de pagina 64
1 2 ... 60 61 62 63 64 65 66 67 68 69 70 ... 186 187

Comentarios a estos manuales

Sin comentarios