133
Technical Data
Carbon Dioxide
Dimensions
Dimensions in mm
Operating Environment
Temperature +5 ... +55 °C (+41 ... +131 °F)
Humidity 0 ... 99 %RH non-condensing
Pressure 700 ... 1200 hPa
Gas flow
operating range < 10 l/min
recommended range 0.2 ... 0.8 l/min
Electromagnetic compatibility Applicable parts of EN61326-1,
Generic Environment
Performance
CO
2
measurement range 0 ... 5 %, 0 ... 10 % or 0 ... 20 %
Accuracy (including repeatability,
non-linearity and calibration
uncertainty) ±(1.5% of range + 3 % of reading)
Long-term stability
0 ... 8 %CO
2
±0.5 %CO
2
/year
8 ... 12 %CO
2
±1 %CO
2
/year
12 ... 20 %CO
2
±2 %CO
2
/year
Response time T
90
< 1 min, when flow > 0.2 l/min
Flow rate dependence
< 1 l/min flow no effect
1 ... 10 l/min flow 4 % of reading/ l/min
Temperature dependence,
typical
-0.3 % of reading/°C
Pressure dependence, typical +0.15 % of reading/hPa
Warm-up time 1 min, 10 min for full
specifications
TUBE I.D. Ø4
(TUBE O.D. Ø6)
mA
V
0
B
A
0
24V
mA
V
0
B
A
0
24V
25
65 6
30
60
96
Ø3.2
Inputs and Outputs
Outputs 4 ... 20 mA, 0 ... 10 V
RS485, 2-wire, non-isolated
Operating voltage 24 V (±20 %) AC/DC
Power consumption <2 W
Ref. B210566EN-D ©Vaisala 2013
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.
Please contact us at
www.vaisala.com/requestinfo
www.vaisala.com
Scan the code for
more information
Comentarios a estos manuales