Technical Data
Dew Point
Area not recommended for
continuous measurement
Dew point measurement renge lower limit
Dew point temperature (°C)
Temperature of measured gas (°C)
Measured Parameters
DEW POINT
Sensor Vaisala DRYCAP
®
180M
Measurement range -60 ... +80 °C (-76 ... +176 °F) Td
For continuous use -60 ... +45 °C (-76 ... +113 °F) Td
Accuracy
up to 20 bar/290 psia ±2 °C/±3.6 °F
(see the accuracy graph below)
20 ... 50 bar/290 ... 725 psia additional inaccuracy +1 °C Td
Dew point accuracy vs. measurement conditions
Response time 63% [90%] at +20 °C gas temperature
Flow rate 1 l/min and 1 bar pressure
-60 ... -20 °C Td (-76 ... -4 °F Td) 5 s [10 s]
-20 ... -60 °C Td (-4 ... -76 °F Td) 45 s [10 min]
TEMPERATURE
Measurement range 0 ... +80 °C (+32 ... +176 °F)
Accuracy ±0.2 °C at room temperature
Temperature sensor Pt100 RTD Class F0.1 IEC 60751
Operating Environment
Operating temperature
for probes -40 ... +80 °C (-40 ... +176 °F)
Mechanical durability up to +180 °C (+356 °F)
of transmitter body -40 ... +60 °C (-40 ... +140 °F)
with display 0 ... +60 °C (+32 ... +140 °F)
Storage temperature range -55 ... +80 °C (-67 ... +176 °F)
Pressure range for probes see probe specifications
Sample flow rate no effect
Measured gases non-corrosive gases
Electromagnetic compatibility Complies with EMC standard
EN61326-1, Industrial environment
Note: Transmitter with display test impedance of
40 ohm is used in IEC61000-4-5 (Surge immunity)
RELATIVE HUMIDITY
Measurement range 0 ... 70 %RH
Accuracy (RH <10 %RH, at + 20 °C) ±0.004 %RH + 20% of reading
PPM
Measurement range (typical) 10 ... 2500 ppm
Accuracy (at + 20 °C, 1 bar) 1 ppm + 20% of reading
Other measurement parameters available (model-dependent):
mixing ratio, absolute humidity, pressure dew point calculated
to 1 bar, temperature difference (T-Td), water vapor pressure
Digital outputs RS232, RS485 (optional)
Service connection RS232, USB
Relay outputs 0.5 A, 250 VAC, SPDT (optional)
Ethernet interface (optional)
Supported standards 10/100Base-T
Connector RJ45
Protocols Telnet
WLAN interface (optional)
Supported standards 802.11b
Antenna connector type RP-SMA
Protocols Telnet
Security WEP 64/128, WPA
Authentication / Encryption
Open / no encryption
Open / WEP
WPA Pre-shared key / TKIP
WPA Pre-shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock
Logged parameters max. three with trend/min./max. values
Logging interval 10 sec. (fixed)
Max. logging period 4 years, 5 months
Logged points 13.7 million points per parameter
Battery lifetime min. 5 years
Display LCD with backlight, graphical trend
display of any parameter
Menu languages English, Chinese, Finnish, French, German,
Japanese, Russian, Spanish, Swedish
Inputs and Outputs
Operating voltage 10 ... 35 VDC, 24 VAC ±20 %
with optional power supply module 100 ... 240 VAC 50/60 Hz
Power consumption @ 20 °C (U
in
24VDC)
RS232 max. 25 mA
U
out
2 x 0...1V / 0...5V / 0...10V max. 25 mA
I
out
2 x 0...20 mA max. 60 mA
display and backlight + 20 mA
during sensor purge max. + 110 mA
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V
Accuracy of analog outputs at 20 °C 0.05% full scale
Temperature dependence of the
analog outputs ± 0.005%/°C full scale
External loads
current outputs R
L
< 500 ohm
0 ... 1V output R
L
> 2 kohm
0 ... 5V and 0 ... 10V outputs R
L
> 10 kohm
Max. wire size 0.5 mm
2
(AWG 20) stranded wires recommended