Vaisala INTERCAP HMDW80 SERIES Especificaciones Pagina 61

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 187
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 60
59
Technical Data
Please contact us at
www.vaisala.com/requestinfo
www.vaisala.com
Scan the code for
more information
Humidity
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Accuracy (incl. non-linearity, hysteresis and repeatability)
temperature range 0 ... +40 °C
0 ... 90 %RH ±1.5 %RH
90 ... 100 %RH ±2.5 %RH
temperature range -40 ... 0 °C, +40 ... +80 °C
0 ... 90 %RH ±3.0 %RH
90 ... 100 %RH ±4.0 %RH
Factory calibration uncertainty (+20 °C)
0 ... 90 %RH ±1.1 %RH
90 ... 100 %RH ±1.8 %RH
Humidity sensor Vaisala HUMICAP
®
180R
Stability ±2 %RH over 2 years
TEMPERATURE
Measurement range -40 ... +80 °C
Accuracy over temperature range
0 ... +40 °C, ±0.2 °C
-40 ... 0 °C, +40 ... +80 °C ±0.4 °C
Temperature sensor Pt1000 RTD Class F0.1 IEC 60751
DEW POINT
Measurement range -40 ... +80 °C
Accuracy (incl. non-linearity, hysteresis and repeatability)
temperature range 0 ... +40 °C
when dew point depression < 15 °C ±1 °C
when dew point depression 15 ... 25 °C ±2 °C
temperature range -40 ... 0 °C, +40 ... +80 °C
when dew point depression < 15 °C - dew point ±2 °C
depression = ambient temperature - dew point
Mechanics
Materials
body stainless steel (AISI 316)
grid filter chrome coated ABS plastic
cable polyurethane or FEP
Housing classification IP65
Body thread MI2x1 / 10 mm
Cable connector 4-pin M8 (IEC 60947-5-2)
Weight
probe 17 g
probe with 0.3 m cable 28 g
Options and Accessories
Sensor protection
plastic grid DRW010522SP
membrane filter DRW010525SP
stainless steel sintered filter HM46670SP
4 ... 20 mA loop power converter UI-CONVERTER-1CB
Mounting bracket for converter 225979
Plastic M12 installation nuts, pair 18350SP
USB cable for PC connection 219690
Probe mounting clamp set, 10 pcs 226067
Probe mounting flange 226061
Connection cables
standard 0.3 m HMP50Z032SP
standard 3 m HMP50Z300SP
80 °C 1.5 m 225777SP
80 °C 3 m 225229SP
180 °C 3 m FEP 226902SP
Inputs and Outputs
Operating voltage
(Use lowest available operating
voltage to minimize heating)
5 ... 28 VDC / 8 ... 28 VDC with
5 V output
8 ... 28 VDC with loop power
converter
Current consumption 1 mA average, max. peak 5 mA
Start-up time
probes with analog output 4 s at operating voltage
13.5 ... 16.5 VDC
2 s at other valid operating voltages
probes with digital output 1 s
Outputs
2 channels 0 ... 1 VDC / 0 ... 2.5 VDC / 0 ... 5 VDC / 1 ... 5 VDC
1-channel loop-power converter (separate module,
compatible with humidity accuracy only) 4 ... 20 mA
digital output (optional) RS485 2-wire half duplex
Operating Environment
Operating temperature range -40 ... +80 °C
Electromagnetic compatibility EN 61326-1: Electrical equipment
for measurement, control and laboratory use – EMC
requirements – for use in industrial locations.
External loads
0 ... 1 V R
L
min 10 kΩ
0 ... 2.5 V /0 ... 5 V R
L
min 50 kΩ
Dimensions
Dimensions in mm (inches)
12
(0.47)
71
(2.8)
M
12
x1
Ref. B210852EN-E ©Vaisala 2013
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifications — technical included — are subject
to change without notice.