Humidity
The Vaisala HUMICAP
®
Humidity and Temperature
Transmitter HMT338 is designed for pressurized processes.
Insert or Remove the Probe while the
Process is Running
With “hot tapping”, the probe is inserted directly into the
process while it is running, without the need for venting or
lowering the process pressure.
The probe is tightened to a ball-valve assembly fixed to
the process pipe or wall. The adjustable hex nut is hand-
tightened to temporarily hold the probe in place. The probe
is then pushed down to the appropriate depth. The hex nut
is then tightened with a wrench to lock the probe in place.
Hot tapping is possible in pressures up to 10 bar.
Typical Applications
process lines
environmental chambers
vacuum-drying processes
compressed air lines with refrigerant dryers
The HMT338 is ideal for installations in pressurized processes
where the probe needs to be removed while the process is
running.
HMT338 Humidity and Temperature Transmitter for Pressurized Pipelines
Technical Data
Temperature measurement range -70 ... +180 °C (-94 ... +356 °F)
Operating pressure 0 ... 4MPa (0 ... 40 bar)
Accessories
Ball-valve set BALLVALVE-1
Pressure fitting ISO 1/2 to NPT 1/2 210662
USB service port cable with PC software 219916
Connection cable for HM70 211339
Wall-mounting plate (plastic) 214829
Pole installation kit with rain shield 215109
DIN rail installation set 215094
PPS plastic grid filter with stainless steel net DRW010281SP
PPS plastic grid filter DRW010276SP
Stainless steel sintered filter HM47280SP
Stainless steel grid filter HM47453SP