Vaisala INTERCAP HMDW80 SERIES Especificaciones Pagina 139

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 187
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 138
137
Spare Parts and Accessories
CO
2
module GM10SP
Temperature Module (CO
2
+T models) TM10SP
Humidity and Temperature Module
(CO
2
+T+RH models) HTM10SP
Decorative cover set (10 pcs.) 236285
Connection cable for HM70 hand-held meter 219980
USB cable for PC connection 219690
Inputs and Outputs
Supply voltage 18 … 35 VDC, 24 VAC ± 20% 50/60 Hz
Current output models
Outputs 0/4…20 mA, 2 and 3 channel models available
Loop resistance 0…600 Ω
Power consumption <2W
Voltage output models
Outputs 0…5/10 V, 2 and 3 channel models available
Load resistance 10 kΩ min.
Power consumption <1W
Digital models
Power consumption <1.5W
Output type RS-485 (galvanic isolation, 1.5 kV)
RS-485 end of line termination Enable with jumper, 120 Ω
Supported protocols Selectable by DIP switch
BACnet MS/TP
Operating mode Selectable Master/Slave
Address range, master mode
Address range, slave mode
0 … 127
128…255
Modbus RTU
Address range 0 … 247
Service port RS-485 line for temporary service use
Mechanics
IP class IP30
Standard housing color White (RAL9003*)
Housing material ABS/PC, UL-V0 approved
Output connector Screw terminals
max. wire size 2 mm
2
(AWG14)
Service port connector 4-pin M8
Weight 163 g
*RAL code is only indicative with potential small variations in color shade
Models
GMW93 CO
2
+T 3-wire, voltage output
GMW93D CO
2
+T 3-wire, voltage output with display
GMW94 CO
2
+T 3-wire, current output
GMW94D CO
2
+T 3-wire, current output with display
GMW93R CO
2
+T+RH 3-wire, voltage output
GMW93RD CO
2
+T+RH 3-wire, voltage output with display
GMW93RA CO
2
+T+RH 3-wire, voltage output with display and
CO
2
indicator LEDs
GMW94R CO
2
+T+RH 3-wire, current output
GMW94RD CO
2
+T+RH 3-wire, current output with display
GMW95 CO
2
+T Digital (BACnet/Modbus) model
GMW95D CO
2
+T Digital (BACnet/Modbus) model with display
GMW95R CO
2
+T+RH Digital (BACnet/Modbus) model
GMW95RD CO
2
+T+RH Digital (BACnet/Modbus) model with
display
GMW90 CO
2
+T Configurable analog/digital models
GMW90R CO
2
+T+RH Configurable analog/digital models
Performance
CARBON DIOXIDE
Measurement range 0 … 5000 ppm
Accuracy
+20 ... +30 °C (+ 68 ... + 86 °F) ± (30 ppm + 2 % of reading)
+10 ... +20 °C, +30 ... +40 °C
(+50 ... +68 °F, +86 ... +104 °F)
± (35 ppm + 2.7 % of reading)
-5 ... +10 °C, +40 ... +55 °C
(+23 ... +50 °F, +104 ... +131 °F)
± (45 ppm + 3.8 % of reading)
Stability in typical HVAC applications Total accuracy at room
temperature ±75 ppm at
600 and 1000 ppm incl. 5 years drift*
Carbon dioxide sensor Vaisala CARBOCAP
®
GM10
TEMPERATURE
Measurement range -5 ... +55 °C (+23 ... +131 °F)
Accuracy
+20 ... +30 °C (+68 ... +86 °F) ±0.5 °C (± 0.9 °F)
+10 ... +20 °C, +30 ... +40 °C
(+50 ... +68 °F, +86 ... +104 °F)
±0.6 °C (± 1.08 °F)
-5 ... +10 °C, +40...+55 °C
(+23 ... +50 °F, +104 ... +131 °F)
±0.8 °C (± 1.44 °F)
Temperature sensor Digital temperature sensor
RELATIVE HUMIDITY
Measurement range 0 ... 95 %RH
Accuracy
Temperature range +10 ... +40 °C (+50 ... +104 °F)
0 ... 60 %RH ±2.5 %RH
60 … 80 %RH ±3.0 %RH
80 ... 95 %RH ±4.0 %RH
Temperature range -5 ... +10 °C, +40 ... + 55 °C
(+23 ... +50 °F, +104 ... +131 °F)
0 ... 60 %RH ±3.5 %RH
60 … 80 %RH ±4.0 %RH
80 ... 95 %RH ±5.0 %RH
Operating Environment
Operating temperature range -5 ... +55 °C (+23 ... +131 °F)
Operating humidity range 0 … 95 %RH
Dewpoint <30 °C (+86 °F)
Storage temperature range -30 ... +60 °C (-22 ... +140 °F)
Electromagnetic compliance EN61326-1, Industrial Environment
Stability in typical HVAC applications ±0.5 %RH/year
Humidity sensor Vaisala HUMICAP
®
180R
*Complies with CEC-400-2008-001-CMF
Technical Data
Carbon Dioxide
Vista de pagina 138
1 2 ... 134 135 136 137 138 139 140 141 142 143 144 ... 186 187