Chapter 5 ________________________________________________________________ Operation
VAISALA______________________________________________________________________ 129
Calibrating Sensors
For calibrating the sensors with MAWS Terminal, the Sensor
Calibration tab is available in the MAWS Stations Settings window,
see Figure 116 on page 129. The list of the MAWS sensors is displayed
in the tab.
0312-038
Figure 116 MAWS Station Settings Window: Sensor Calibration
Tab
NOTE
You must load an appropriate setup to MAWS before you are able to
use the Sensor Calibration tab.
NOTE
The Sensor Calibration tab is not visible for setups made with
MAWS software versions prior to 3.06. With old setups, MAWS
common parameters frame only shows the calibration values for the
wind direction and solar radiation sensors, and other sensors need to
be calibrated through the terminal connection with the appropriate
calibration commands.
CAUTION
When you upgrade a previous MAWS version to 3.06 or above, the
following sensors will not operate correctly in the Sensor Calibration
tab until they are removed from the setup and then re-created: CM6B,
CM11, QMS101, QMS102, and QMN101.