Chapter 4 _______________________________________________________________ Installation
VAISALA_______________________________________________________________________ 87
Drilling a Hole for the Sensor
Drill a hole according to the following procedure:
1. Choose a desired location for the sensor. Assure that the hole is
located within cable length of the logger enclosure.
2. Drill a hole into the ground with the auger held as straight as
possible. After you have drilled about 0.2 m (8 in.), extract the
auger from the hole.
0106-039
Figure 75 Drilling Procedure
3. Remove soil from the auger, for example, with a screwdriver not
with fingers. Refer to Figure 76 on page 88.
CAUTION
The installation hole must be straight. Otherwise the sensor may bend,
and the resulting tension damages the sensor.
CAUTION
Do not use a hammer to pound the auger into ground even if the ground
is hard or especially when it is rocky.
NOTE
With fluid or loose soils there is risk of soil collapsing into the hole.
To prevent this you can use a plastic or metal tube to assist the
installation. After you have installed the sensor, remove the tube. Fill
the gap with sand, mould, or other fine-grained soil with no sharp
stones.