Chapter 8 ____________________________________________________________ Technical Data
VAISALA______________________________________________________________________ 225
If the primary cell setting is detected, the charger circuitry of the logger
is completely shut off to protect the battery from reverse currents.
Lead Battery Charger Operation
The charger has three operating modes when a lead battery is detected:
Normal Charge, Quick Charge and Float Charge. When the logger
starts, it first checks the battery type, and if lead is found, the charging
task starts and is performed once a minute.
Normal Charging
At first, the charging task always enters Normal Charging mode. In
Normal Charge, charging voltage is set to 6.85 V corrected with
temperature coefficient. Charging current is set to 300 mA, 500 mA, or
700 mA depending on the battery capacity.
If the battery is accepting certain amount of charging current and there
is enough energy available at +ExtDC input, the charging task enters
Quick Charge mode. In the opposite case, when the charging current
drops below 0.0075 CA, even if more energy would be available, the
charging task enters Float Charge mode.
If the battery is discharging (charging current is negative), the charging
task always enters Normal Charge mode.
NOTE
Short pins 3 and 4 in the internal battery connector with jumper or
preferably use a special pin header for correct battery type sensing.
CAUTION
Never connect voltages higher than 10 V between the terminals
+BATT and GND as this may damage the logger electronics.
NOTE
If the internal temperature of the logger is below -20 °C (-4 °F), the
charging current is limited to 100 mA regardless of the battery
capacity.
NOTE
After powering up or reset, it takes several minutes to calculate the
remaining capacity of the battery if the charger remains in Normal
Charge mode.